HB103 Medical Infant Incubator Baby Caring Equipment In Hospital
HB103 Medical Infant Incubator Baby Caring Equipment In Hospital

HB103 Medical Infant Incubator Baby Caring Equipment In Hospital HB103 Medical Infant Incubator Baby Caring Equipment In Hospital

Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.

Quantity:
Share this:
  • Product Description
  • Parameter

HB103 Hospital Infant Incubator Features:
1.Servo Control Mode: air temperature, skin temperature;
2.Instant Parameter Displaying: set temp, air temp ,skin temp, internal humidity, heat power rate
3.Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure;
4.>37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
5.Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
6.RS-232 Output Connector (Use for network system ).
7.Export model: install the guardrail instead of top shelf.
Neonatal Baby Warmer, Hospital Infant Incubator, Medical Infant Warmer, Baby Transport Incubator

HB103 
Mattress to Hood450mm
Observation side of hoodTwo incline
Operate port6
Front access panel826*310mm
Mattress tray size630*360mm
Power supply≤400VA
Temperature controlled by microcomputerAir and skin temp
Air temp
Air temp control range20℃-37℃
37℃-39℃
Skin Temp control range30℃-37℃
37℃-38℃
Skin Temp sensor accuracy≤0.3℃
Humidity display range5%-99%
Temperature fluctuation≤0.5℃
Mattress temperature uniformity≤0.8℃
Temperature rise time≤30min
Noise level within hood environment≤55dB
Alarms 
Power failure alarmyes
Fan failure alarmyes
Sensor failure alarmyes
High air temp alarm+1.5℃
Low air tem alarm-2.5℃
High skin temp alarm+1.0℃
Low skin tem alarm-1.0℃
Over air tem alarm≤38℃, ≤40℃
Over skin tem alarm≤38℃, ≤39℃
System failure alarmyes
Alarm silence functionyes
Operational condition 
Environment temperature+20℃-+30℃
Environment relative humidity5%-99%
Environment air velocity of flow<0.3m/s
Others 
RS232 interfaceyes
Railingyes

Infant bed tilt adjustment

±10° step less

Max load of mattress tray

10kg

IV Pole Max load

2kg

×