HB-YP930 Hospital Medical Equipment Premature Baby Care Infant Incubator

HB-YP930 Hospital Medical Equipment Premature Baby Care Infant Incubator HB-YP930 Hospital Medical Equipment Premature Baby Care Infant Incubator

Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.

Quantity:
Share this:
  • Product Description

Brief Introduction:
Two control modes: Air mode and baby mode controlled by micro-computer; Humidity control system;
Set temperature, air temperature, baby temperature, humidity and oxygen are displayed separately;
>37℃ temperature set function;
Failure alarm indication;
Embedded integrated sensor box;
Integrated water reservoir in the shape of drawer;
Double wall hood, automatic air circulation device;
Inclination of bassinet is adjustable;
RS-232 connector;
DC motor.
Standard Configuration:
Optional Configuration:
Fixed neonate bilirubin phototherapy equipment model: HBXHZ-90 (with blue fluorescent lamps)
Fixed neonate bilirubin phototherapy equipment model: HBXHZ-90L (with LED blue lights)
Vertical height adjustment stand (VHA stand) Weighing system, Disposable skin temperature sensor and SP02.
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 1000VA
Mode of control: Air mode and baby mode controlled by micro-computer
Air temperature control range: 25℃ -37℃ (override mode 37-39℃ )
Baby temperature control range: 34-37℃ (override mode 37-38℃ )
Accuracy of skin temperature sensor: ±0.2℃
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition)≤0.5℃
Temperature uniformity: 0.8℃ (mattress is on the horizontal position)
Inclination of the bassinet: ±12°
Noise inside hood: 45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, sensor box position alarm , low water alarm , tank position alarm and system alarm.
Humidity display range: 0%RH-99%RH
Humidity control range: 0%RH-90%RH
Accuracy of humidity display: ±5%RH(environment temperature is 25℃ , environment humidity is 45%RH±5%RH)
Indication range of oxygen concentration: 0%-99%
Control range of oxygen concentration: 20%-60% ( for HB-YP970 )
Accuracy of oxygen display concentration: ±2%(oxygen concentration sets 25%) ±3%(oxygen concentration sets >25%)
Environmental temperature:
Operating range: 20℃ -30℃ (Set temperature should be higher 3℃ than the environment temperature)
Ambient air movement rate: <0.3m/s
Product Packaging:
Package:each unit is packed into two cartons
Dimension (Main body): 1085mm x 640mm x 915mm
                  (Cabinet): 1125mm x 665mm x 730mm
Gross Weight (Main body):66Kg 
                       (Cabinet): 60Kg

×