Baby mode control by micro-computer;
Failure alarm indication;
Blue light radiation in two sides;
Timer for phototherapy caring time;
Bassinet can be pulled out from the side door.
DC motor
HBXHZ Hospital Medical Baby Neonate Bilirubin Phototherapy with Side Door
Brief Introduction
Baby mode control by micro-computer;
Failure alarm indication;
Blue light radiation in two sides;
Timer for phototherapy caring time;
Bassinet can be pulled out from the side door.
DC motor
Standard configuration
Above case (including above lamp, controller and infant bed);
Nether case (including nether lamp and cabinet);Skin temperature sensor;I.V.Pole.
Optional Configuration:
LED blue light sa upper case light source;
Disposable skin temperature sensor is optional.
Specification
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 600VA
Mode of control: Air mode control by micro-computer
Temperature control range: 25-34°C
Skin temperature sensor display range: 5°C -65°C
Correlation of displayed average value and actual average value ( steady temperature condition )≤0.8°C
Accuracy of skin temperature sensor: ±0.3°C
Temperature uniformity of the bed≤ 2.0°C
Noise level above the bed: 55dB(A)
Failure alarm: over-temperature alarm; deviation alarm, sensor failure alarm, fan failure alarm and check alarm.
The total irradiance for bilirubin on the effective surface area:
Above case≥1.5mw/cm2 (blue fluorescent lamp)
≥ 2.0 mw/cm2(LED blue light)
Nether case≥3.0 mw/cm2
Average value of general bilirubin radiation on the effective surface area:
Above case≥1.4mw/cm2 (blue fluorescent lamp)
≥1.5mw/cm2 (LED blue light)
Nether case: 2.5mw/cm2
Uniformity of the total irradiance for bilirubin: >0.4
Dominant wave length of blue light: 400nm-550nm
Environmental temperature:
Operating range : 18 °C -28°C
Ambient air movement rate:<0.3m/s
Product Packaging:
Package: each unit is packed into two cartons
Dimension (Main body): 1000mm x 700mm x 805mm
(Cabinet): 900mm x 600mm x 885mm
Gross Weight (Main body): 70Kg
(Cabinet): 50Kg