HBNBB-Ⅳ Baby Neonate Bilirubin Phototherapy Brief Introduction:
Intensive phototherapy equipment, large effective treatment area, meets AAP Guidelines for intensive phototherapy;
Double side LED phototherapy radiation;
Bed temperature mode servo controlled by micro-computer;
Baby temperature detection, LCD color display;
Self-checking and multiple failure alarm;
Timer for the phototherapy time;
Height of upper light box is adjustable;
Environment temperature detection.
Standard Configuration:
Upper light box (including operation screen), bed, mattress, guard, storage box and mobile stand.
Specification:
Power supply: AC220V-230V/50Hz or AC110-120V/50-60Hz, 150VA
Temperature control mode: bed temperature control mode
Bed temperature control range: 35°C -38°C
Skin temperature display range: 5°C -65°C
Accuracy of bed temperature control≤±1.0°C
Accuracy of bed temperature display≤±0.7°C
Bed temperature variability≤: ±0.5°C
Heating up time: 2 hours (the mattress temperature rise from23°C to 37°C ) Accuracy of skin temperature sensor≤ ±0.2°C
Mattress temperature uniformity≤±2.5°C
Working noise of bed surface: 55dB(A)
Failure alarm: over-temperature alarm, sensor alarm, temperature deviation alarm, motor fan failed alarm
General bilirubin radiation of the effective surface area:
Upper phototherapy:
Irradiation distance cm | LED | Average value of general bilirubin radiation mW/cm2 | General radiation M WC M 2 | Maximum value of general bilirubin radiation mW/cm 2 |
28 | ALL ON | ≥3.1 | ≥3.5 | ≥4.5 |
38 | ALL ON | ≥2.6 | ≥3.0 | ≥4.2 |