Pneumatic Driven Ventilator SIMV Mobile Anesthesia Ventilator Features:
1.Applied for adult and pediatric
2.Microprocessor controlled ,integrated anesthetic ventilation system
3.7 inch TFT display screen
4.Good air tightness, high reliability

Pneumatic Driven Ventilator SIMV Mobile Anesthesia Ventilator Features:
1.Applied for adult and pediatric
2.Microprocessor controlled ,integrated anesthetic ventilation system
3.7 inch TFT display screen
4.Good air tightness, high reliability
|
Technique Specifications
|
|
|
Dimensions: |
43cm*38cm*49cm |
|
Weight: |
10kg |
|
Working type: |
Electronically controlled, pneumatic driven |
|
Patient type: |
Pediatric, Adult |
|
Ventilation mode: |
VCV, SIMV, PCV, MANUAL(BAG) |
|
Tidal volume: |
20-1500ml |
|
l:E ratio: |
4:1-1:8 |
|
Frequency: |
|
|
Adjustment range: |
1-100bpm |
|
Under IPPV and Manual mode: |
4~65bpm |
|
Under SIMV: |
1-40bpm |
|
Monitoring: |
VT, MV, BPM, Paw, FiO2, PEEP, Respiratory Frequency, l:E ratio, Ppeak |
|
Waveform: |
Pressure- time, Flow- time, F-V loop, P-V loop |
|
Power: |
AC 220V±10%, 50HZ±2%, built-in battery > 2hrs(rechargeable) 12V |
|
Inspiratory time (Tl) (at SIMV) |
Adjustment range:0.2 s 〜6.0 s (the limit of l:E ratio< 4) |
|
Pressure trigger: |
—20 cmH20 〜+ 20 cmH20 |
|
Pressure limit: |
|
|
Patient airway pressure range: |
-2KPa 〜10KPa |
|
Airway pressure monitoring range: |
0〜10KPa |
|
PEEP: |
3 ~ 20cmH2O |
|
Safety valve pressure limit: |
< 6Kpa |
|
Alarm: |
High(Low) MV alarm,AC power lost alarm. |
|
Continuous positive airway pressure, Apnea, Self-test failure, Battery Low |
|
|
High(Low)Respiratory Frequency, High(Low) FiO2 |
|
|
Accessories: Oxygen sensor Silicon tubes (reusable) Tidal volume sampling sensor Oxygen high pressure hose
|
|
|
Oxygen sensor
|
|
|
Silicon tubes (reusable)
|
|
|
Tidal volume sampling sensor
|
|
|
Oxygen high pressure hose |
|