AKL600-II Hospital Portable Ceiling LED Light Operating Room Lamp With Shadowless Lighting

AKL600-II series shadow less operation lamp is suitable for the requirements of various operation conditions and it is ideal operating lighting equipment for Modelrn operation rooms and clean operation rooms.
Get a Quete

AKL600-II Hospital Portable Ceiling LED Light Operating Room Lamp With Shadowless Lighting

1579076133734806.jpg

AKL600-II series shadow less operation lamp is suitable for the requirements of various operation conditions and it is ideal operating lighting equipment for Modelrn operation rooms and clean operation rooms.

Features:

The no welding titanium alloy main arm is adopted, so that the appearance is aesthetic and the rotation is more flexible and convenient.

The removable handle sheath can be disinfected at the high temperature of 135℃.

The demolished panel made of imported material makes the light rays softer.

The imported switch power supply is adopted to control the voltage, making the work voltage permanently stable.

The special radiation filter reducing the heat to the minimum level.

The multi-flake retro reflector made of imported aluminum sheet has the features of no coating film and never fading.

Imported lamp receptacle, Tungsten halogen bulbs of international brand are adopted with the average service more than 1500 hours.

Light adjustment circuit adopts CPU control, enabling increase or decrease of illumination and boasting brightness memory.

One lamp holder with two bulbs enables automatic operation of the standby bulb in 1/2 second when the main bulb malfunctions, and it has indicators for the damage of main or standby bulb.

 

医疗细节(合并版1).jpg


 

Model                                    AKL600/600-II

Illumination(lux)                           140000≥Ec≥80000

140000≥Ec≥80000

Color Temperature(K)                       4500±500

Color rendering index                       100≥Ra≥85

Diameter of Spot(mm)                      d10175±35

Depth of light beam(mm)                    ≥700

Brightness level adjustment                  CPU light-dimmer

The value of the total irradiance               <6m W/m²lx

(service range for temperature rise≤10℃)

Total irradiance(Temperature rise on the head   ≤600 W/m²

Of patient under surgical operation≤2℃)

Diameter of  lamp holder(mm)                600/600

Shift of major and stand by lamps              Automatic

Bulb power W(24V)                          150/150

Power supply(AC)                           220±20% 50Hz

Input power(VA)                          440+10%

Minimum height for installation(mm)           2800~3200


Model                                    AKL600-II  or KL600-II(Wall)

Illumination(lux)                           140000≥Ec≥80000