HB106 Infant Radiant Warmer Neonatal Incubator Features:
1.Temperature Control Mode: pre-heating control, servo control & manual control;
2.Instant Parameter Display: set temp, skin temp, heat power rate;
3.Built-in APGAR Timer and >37°C preset function
4.Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
5.Tilting Bed with Rotary Radiant Head;
6.X-ray Cassette Tray under infant bed, examination light of high illuminance;
7.Use anti-blast quartz infrared tube as radiant heating source;
8.With phototherapy unit;
9.With two oxygen cylinders;
10.With Suction, with neonate resuscitation.
Specifications:
1.Medical infant radiant warmer
2.Transparent protector
3. Alarms are audible and visual

HB106 Medical Neonatal Radiating Warmer Newborn Warmer Tables Infant Radiant Warmer
Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. ... A heat shield of thin polyethylene film can be used with a radiant warmer to reduce heat loss by convection and evaporation.
Get a Quete
| Operational condition | +18℃-+30℃ |
| Environment temperature | 30%-75% |
| Environment relative humidity | <0.3m/s |
| voltage | ≤1000VA |
| Temperature control mode | Pre-heat, auto and manual |
| Temperature control range | +25℃-+37℃ |
| +37℃-+38℃ | |
| Temperature sensor uniformity | ≤0.3℃ |
| Mattress temperature uniformity | ≤2℃ |
| Temperature rise time | ≤45min |
| Heat output indicate | 0-100% |
| Radiant box horizon angle adjust range | 0±90° step less |
| Mattress tray size | 68*56cm |
| Mattress tilt adjust range | 0±10° step less |
| Mattress tray Max.Load | 10kg |
| Transfusion pole Max. Load | 2kg |
| Shelf Max.Load | 2kg |
| X-ray shelf | yes |
| RS-232 connector | yes |
| Mattress to Radiant box | 81cm |
| Mattress infrared wave length | 1-3um |
| APGAR timer | yes |
| Power failure alarm | yes |
| Sensor failure alarm | yes |
| Temperature deviation alarm | ±1℃ |
| Over temperature alarm | ≤38℃, ≤39℃ |
| Low-pressure suction | yes |
| Operating noise | ≤55db |
| White phototherapy light | 1set |
| 7L oxygen cylinder(with oxygen manometer and oxygen flow meter) | 2sets |
| Neonate resuscitation bag | 1set |
Request More Information
Related Products
HB104 Medical Hospital Infant Care Baby Radiant Warmer
HB102 Hospital Care Equipment Newborn Baby Infant Radiant Warmer
HBHKN-2001 Medical Hospital Clinic Baby Care Infant Neonate Radiant Wamer
HBHKN-9010 Medical Hospital Clinic Baby Care Infant Neonatal Radiant Wamer
HBHKN-2000 Medical Hospital Clinic Baby Care Infant Newborn Radiant Wamer
HBHKN-90 Medical Hospital Clinic Newborn Infant Baby Care Radiant Warmer
HBHKN-93/ HBHKN-93A / HBHKN-93B Hospital Medical Baby Care Newborn Infant Radiant Warmer