I. Technical parameters
1. Electric parameters
1.1 Anode voltage: 60~92 KV
1.2 Anode current: 1~15 mA
1.3 Max. output power: 1.38 KW
1.4 Parameter adjustment range: 60~92 KV, 1~15 mA
1.5 Information display: KV, mA, information of up/down position, characteristics of human body, laser designation and locating
2. X-ray tube assembly
2.1 Tube model: D-054SB (Toshiba)
2.2 Focal spot: 0.5mm
2.3 Thermal capacity: 35KJ (50KHU)
2.4 Radiography mode: CBCT
3. Imaging system (THALES flat panel detector 650HD-E)
3.1 Image receiving size: 12cm X 15cm
3.2 Pixel size: 150 um
3.3 Valid field of view: 80 (diameter) X 90 (height)
3.4 Display resolution: 1280 x 1024
3.5 Rotary angle: 200°
3.6 Rotary time: 18 S
3.7 Exposure time: 4~8 S
3.8 Scan number: 400 frames
3.9 Reconstruction time: ≤18 S
3.10 Voxel size: 0.2 mm
3.11 A/D conversion: 14 bits
3.12 Acquisition mode: pulse mode
3.13 Detector type: CMOS
3.14 Pixel matrix: 960 x 786
3.15 Resolution: 2D ≥ 3.1 lp/mm
4. Structural performance
4.1 Stand column up/down range: 1000 mm
4.2 Noise:<70dB
4.3 Anti-collision protection: YES
4.4 Complete machine weight: 260 Kg
4.5 Dimension: 788 mm (W) x 1090 mm (L) x 2200 mm (H)
4.6 Power requirement: 220 V/50 Hz/2 kVA
4.7 Space requirement: 5 sqm, single side 2 m
4.8 Radiology room should be of at least 2 mm lead equivalent
4.9 Electrical control device size: 340 mm (W) x 550 mm (D) x 650 mm (H)
II. Standard configuration
1. Rack 1 set
2. High frequency generator 1 set
3. 19” color general LCD, 24” color general LCD or
21” medical grade LCD (choose 1) 1 pc
4. 12 cm x 15 cm flat panel detector 1 set
5. Digital acquisition workstation 1 set
6. Electrical adjustable multi leaf collimator 1 set
7. Graphical LCD touch screen 1 pc
8. Wall mounted control box 1 pc
9. Electrical control cabinet 1 unit
10. Printer (Optional) 1 unit