WRSP-605T Target Controlled Infusion Pump TCI Electric Syringe Pump WRSP-605T Target Controlled Infusion Pump TCI Electric Syringe Pump

1. Applicable Syringe: 10, 20, 30, 50/60 ml 2. 0.01 ml/h Increments for flow rate, bolus rate,KVO rate 3. Wireless Management

Quantity:
Share this:
  • Product Description
  • Parameter

Target Controlled Infusion Pump TCI Electric Syringe Pump Features:

1. Selection of the induction method and setting of the target concentration according to the basic parameters of the patient;

2. Selection of the site of target control, plasma target control or effect chamber target control;

3. Adjusting the target concentration according to the anaesthetic effect

4. Adjusting the dose according to the duration of the operation.

5. Six modes of administration: simple mode, speed mode, time-dose mode, weight mode, plasma target control, effect chamber target control;

6. Three induction modes: rapid induction, step induction and smooth induction, in line with clinical dosing habits;

7. High-resolution colour screen with adjustable backlight;

8. Multi-position adjustable blocking pressure;

9. Keyboard can be locked during operation, real-time display of parameters during operation;

10. Target-controlled syringe pump can be combined with DPS function.




target controlled infusion pump, tci pump, electric syringe pump

target controlled infusion pump, tci pump, electric syringe pump

ModelWRSP-605T
Syringe Size10, 20, 30, 50/60 ml
Applicable SyringeCompatible with syringe of any standard
VTBI1-1000ml( in 0.1, 1, 10 ml increments)
Flow RateSyringe 5 ml: 0.1-100 ml/h ( in 0.1, 1, 10 ml/h increments)
Syringe 10 ml: 0.1-300 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-800 ml/h
Syringe 50/60 ml: 0.1-1200 ml/h
Bolus RateSyringe 5 ml: 0.1-100 ml/h ( in 0.1, 1, 10 ml/h increments)
Syringe 10 ml: 0.1-300 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-800 ml/h
Syringe 50/60 ml: 0.1-1200 ml/h
Anti-BolusAutomatic
Accuracy±2%(mechanical accuracy 1%)
Infusion ModeFlow rate: ml/min, ml/h
Time-based
Body weight
KVO Rate0.1-1 ml/h (in 0.01 ml/h increments)
AlarmsOcclusion, near empty, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off
Additional FeaturesReal-time infused volume, automatic power switching, automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory, history log
Drug LibraryAvailable
Occlusion SensitivityHigh, medium, low
History log50000 events
Power Supply, AC110/230V (optional), 50-60 Hz, 20VA
Battery14.8V, rechargeable
Battery Life8 hours at 5 ml/h
Working Temperature5-40℃
Relative Humidity20-90%
Atmospheric Pressure700-1060hpa
Size245*120*115 mm
Weight2.5kg
Safety ClassificationClass Ⅱ, type BF


×