1. Applicable Syringe: 10, 20, 30, 50/60 ml 2. 0.01 ml/h Increments for flow rate, bolus rate,KVO rate 3. Wireless Management
Target Controlled Infusion Pump TCI Electric Syringe Pump Features:
1. Selection of the induction method and setting of the target concentration according to the basic parameters of the patient;
2. Selection of the site of target control, plasma target control or effect chamber target control;
3. Adjusting the target concentration according to the anaesthetic effect
4. Adjusting the dose according to the duration of the operation.
5. Six modes of administration: simple mode, speed mode, time-dose mode, weight mode, plasma target control, effect chamber target control;
6. Three induction modes: rapid induction, step induction and smooth induction, in line with clinical dosing habits;
7. High-resolution colour screen with adjustable backlight;
8. Multi-position adjustable blocking pressure;
9. Keyboard can be locked during operation, real-time display of parameters during operation;
10. Target-controlled syringe pump can be combined with DPS function.
Model | WRSP-605T |
Syringe Size | 10, 20, 30, 50/60 ml |
Applicable Syringe | Compatible with syringe of any standard |
VTBI | 1-1000ml( in 0.1, 1, 10 ml increments) |
Flow Rate | Syringe 5 ml: 0.1-100 ml/h ( in 0.1, 1, 10 ml/h increments) |
Syringe 10 ml: 0.1-300 ml/h | |
Syringe 20 ml: 0.1-600 ml/h | |
Syringe 30 ml: 0.1-800 ml/h | |
Syringe 50/60 ml: 0.1-1200 ml/h | |
Bolus Rate | Syringe 5 ml: 0.1-100 ml/h ( in 0.1, 1, 10 ml/h increments) |
Syringe 10 ml: 0.1-300 ml/h | |
Syringe 20 ml: 0.1-600 ml/h | |
Syringe 30 ml: 0.1-800 ml/h | |
Syringe 50/60 ml: 0.1-1200 ml/h | |
Anti-Bolus | Automatic |
Accuracy | ±2%(mechanical accuracy 1%) |
Infusion Mode | Flow rate: ml/min, ml/h |
Time-based | |
Body weight | |
KVO Rate | 0.1-1 ml/h (in 0.01 ml/h increments) |
Alarms | Occlusion, near empty, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off |
Additional Features | Real-time infused volume, automatic power switching, automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory, history log |
Drug Library | Available |
Occlusion Sensitivity | High, medium, low |
History log | 50000 events |
Power Supply, AC | 110/230V (optional), 50-60 Hz, 20VA |
Battery | 14.8V, rechargeable |
Battery Life | 8 hours at 5 ml/h |
Working Temperature | 5-40℃ |
Relative Humidity | 20-90% |
Atmospheric Pressure | 700-1060hpa |
Size | 245*120*115 mm |
Weight | 2.5kg |
Safety Classification | Class Ⅱ, type BF |