HB102 Hospital Care Equipment Newborn Baby Infant Radiant Warmer HB102 Hospital Care Equipment Newborn Baby Infant Radiant Warmer

1.Servo Control Mode: air temp, skin temp;
2.Instant Parameter Displaying: set temp, air temp ,skin temp, heater power rate;
3.Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure;

Quantity:
Share this:
  • Product Description

HB102 Hospital Care Equipment Newborn Baby Infant Radiant Warmer

Features:

1.Servo Control Mode: air temp, skin temp;

2.Instant Parameter Displaying: set temp, air temp ,skin temp, heater power rate;

3.Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure;

4.>37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;

5.Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;

6.RS-232 Output Connector (Use for network system ).

7.Export model: install the guardrail instead of top shelf

 


HB102

Mattress to Hood

450mm

Observation side of hood

Two incline

Operate port

6

Front access panel

826*310mm

Mattress tray size

630*360mm

Power supply

≤400VA

Temperature controlled by microcomputer

Air and skin temp

Air temp

Air temp control range

20℃-37℃

37℃-39℃

Skin Temp control range

30℃-37℃

37℃-38℃

Skin Temp sensor accuracy

≤0.3℃

Humidity display range

-------------

Temperature fluctuation

≤0.5℃

Mattress temperature uniformity

≤0.8℃

Temperature rise time

≤30min

Noise level within hood environment

≤55dB

Alarms


Power failure alarm

Yes

Fan failure alarm

Yes

Sensor failure alarm

Yes

High air temp alarm

+1.5℃

Low air tem alarm

-2.5℃

High skin temp alarm

+1.0℃

Low skin tem alarm

-1.0℃

Over air tem alarm

≤38℃, ≤40℃

Over skin tem alarm

≤38℃, ≤39℃

System failure alarm

yes

Alarm silence function

yes

Operational condition


Environment temperature

+20℃-+30℃

Environment relative humidity

5%-99%

Environment air velocity of flow

<0.3m/s

Others


RS232 interface

yes

Railing

yes

            Infant bed tilt adjustment

                                   ±10° step less

           Max load of mattress tray

10kg

                 IV Pole Max load

2kg

 


×